Duniway Stockroom Corp. rebuilds Ion Pumps from many other manufacturers, including:
| Anelva | Cambridge | Cameca | 
| Edwards | JEOL | Kratos | 
| Leybold | Meca | Riber | 
| Satsuki | Thermionics | 
Please inquire about the rebuilding prices for these Ion Pumps or any other types you might have.
Ion Pump Rebuilding Process
- Open Pump & Remove Elements
 - Disassemble Elements Completely
 - Clean Stainless Steel Parts
 - Replace Titanium and Tantalum Parts (Tantalum, if needed, for additional cost)
 - Clean and/or Replace Ceramic Parts
 - Vacuum Degas all Metal Element Parts
 - Reassemble Elements
 - Install the Elements in the Clean Body
 - Leak Check
 - Bake-Out on Vacuum Manifold
 - Check for leakage current
 - Pinch off Seal
 - Clean Exterior
 - Check for Leakage Current