Supplying reliable vacuum equipment since 1976!

Page Index

Get Catalog

 Search Site

  HI-POTTER FOR ION PUMPS
 


When ion pumps develop field emission leakage currents due to build up of deposits with microscopically small, sharp points, the leakage current can be reduced or eliminated by the technique known as “hi-potting”.  This involves application of high voltage, typically 10-15 KV, either DC or AC, to the ion pump high voltage feedthrough.

One supplier of equipment for hi-potting units is:

 HiPotronics

http://www.hipotronics.com/

845-279-8091, 800-727-HIPO, FAX 845-279-2467

One model which we use is:
DC Insulation Tester
Model 220, Type AS11-582-3
(0-20KV, 15MA)
Duniway Stockroom Corp.

 

 


DUNIWAY STOCKROOM CORP
.
1305 Space Park Way w
 Mountain View, CA 94043
Tel: 650-969-8811 w 1-800-446-8811 w Fax: 650-965-0764
w  w   Trademarks® pdf   w  Company Photo
w get Acrobat Reader w

* Prices subject to change; please review Catalog for latest prices

Site by Gryphon Web Design, Palo Alto, Ca
Rev:
April 29, 2008