|
Using the Terranova 741 to Diagnose Field Emission Leakage
One of the many advantages of using sputter-ion pumps, is that the discharge current is approximately
proportional to pressure over wide ranges, from 10-4 torr to 10-10
torr.
However, after extended use and multiple bakeouts,
deposits of sputtered material can build up. Some of these deposits may generate sharp points between the cathode and
anode of the pump. These sharp points,
in the presence of the 5 KV to 7.5 KV operating voltages of ion pumps can give
rise to field emission leakage currents.
The field emission currents can be several microamps
in magnitude, thus masking the true discharge ion current. For example, in a
typical 100 l/s ion pump, a field emission leakage current of 5 microamps would
start to mask pressure readings below 10-8 torr and make readings in
the 10-9 torr range impossible. In addition, in many cases, the
field emission current has a ‘noisy’ appearance with
10-20% variation. The existence of field emission leakage can
be verified by removing the pump magnets, and observing the current. Any current would be due to
non-discharge/pumping mechanisms. However, removal of magnets on larger pumps, while installed on a system
can be awkward.
With most ion
pump control units it is difficult if not impossible to diagnose such leakage
currents. In addition, low voltage
resistance measurements show nothing, because the field emission leakage
currents rise exponentially with voltage, and at V-O-M voltage levels the
current is extremely small. Figure 1
below shows a typical graph of field emission current (lower, exponential
curve) versus voltage. (Actual curves
may be the sum of the linear ion current and the exponential field emission
leakage current, but the exponential component should be obvious.)
|
|
Figure 1
However,
with an ion pump control unit like the Terranova 741, the ion pump operating
voltage can be adjusted during operation. So, when low pressures should have been reached, but the ion pump
current is reading a few microamps, it is straight forward to use the SETUP
function to vary the high voltage from 3.5 to 7.5 KV (steps of 1 KV should tell
the story) and record the current for each voltage setting. If you get a curve with the exponential
shape, like the lower curve, your pump has more than likely developed field
emission leakage.
Field
emission leakage currents do not in any way interfere
with or diminish the ion pump’s ability to pump. They only mask the ability to
use the pump current to indicate the pressure at the bottom of the pressure
range. In smaller pumps, with only a few
anode cells, field emission may actually speed up the striking of the discharge
at UHV pressures, by contributing the initial electron required to start the
avalanche.
This condition can usually be remedied by
“Hi-Potting” the pump, by applying significantly higher voltage (15-20KV) to
the high voltage feedthrough. The high,
exponential currents flowing through the fragile field emission points will
usually burn them off. Sometimes, judicious light tapping with a
non-magnetic object can break the field emission points loose, also. Call Duniway Stockroom Corp. to obtain information on “Hi-Potting”
apparatus.
If, on the other hand, you find a more linear
relationship between applied voltage and current, it is most likely due to real
ions, thus indicating the pressure in the system. If the current and therefore the pressure are
higher than expected, well, you probably have a leak or need more time and
bakeout cycles to get the pressure down to desired values.
In unusual cases, linear current
versus voltage might indicate a resistive load. This could be dirt or fingerprints on the external ceramic portion of
the high voltage feedthrough. In that
case, cleaning with solvent and DI water, with mild heating with a heat gun
will take care of the leakage. If,
however, the resistive coating is on the internal ceramics of the ion pump, a
complete cleaning and processing is required. Such resistive coatings could come from extended operation of the ion
pump at high pressures where the glow-discharge is not confined to the pump’s
magnetic field. Or, the coatings could
come from active metals, such as barium, cesium, etc. which may have been present
in the vacuum system and deposited out on the ion pump insulators.
November,
2001
S.
Rutherford
Duniway
Stockroom Corp.
|